#295 new Reception
12 Apr 17:00 Test of D and C reception lines
Reported by: | imanay | Owned by: | |
---|---|---|---|
Priority: | normal | Milestone: | 2017 JULIA |
Component: | JULIA | Version: | |
Keywords: | Cc: | ||
Parent Tickets: |
Description (last modified by imanay)
Due to 150km echoes at the D channel lower than C, these reception lines were tested with a noise generator (Attached the connections of this test).
For the following cables, these results were obtained at the acquisition system:
- 50 ohmios load instead of noise generator:
C | 7.98 |
D | 8.06 |
- Using noise generator:
C | D |
16.41 (4,4.75)m | 12.00 (4,5.5)m |
17.05 (4,4)m | 16.66 (4,4.75)m |
15.78 (4,4.75)m | 17.48 (4,4)m |
15.17 (4,4.75)m | 12.94 (4,5.5)m |
Subtickets (add)
Attachments (1)
Change History (3)
Changed 8 years ago by imanay
comment:1 Changed 8 years ago by imanay
- Description modified (diff)
comment:2 Changed 8 years ago by imanay
- Description modified (diff)
Note: See
TracTickets for help on using
tickets.
Connection of the C and D reception lines test