Opened 9 years ago

Last modified 9 years ago

#899 new Other Instruments

05 Feb 19:40 VIPER (Ionosonde) and Digisonde (and synchronization)

Reported by: kkuyeng Owned by:
Priority: normal Milestone: 2016_02_1 ISCOD Gravity Waves
Component: ISR+Imaging Version:
Keywords: Cc:

Description

VIPER configuration

  • ionograms every 5 minutes (3, 8, 13, 18, ...), so it won't interfere with digisonde pulses.
  • IPP = 3000km (ISR + Imaging is synchronized with PPS)

Digisonde
-ionogram every 5 min (0,5,10,15,...)

  • its tx pulse is seen at the top of the 1000km of the ISR + Imaging experiment.

*ISR+Imaging

  • syncrhonized with PPS, then must restart Digisonde clock (10MHz) to put its tx pulse as close as the PPS (shown on osciloscope). Right now it is ok, but in case of a blackout this needs to be done.

Change History (1)

comment:1 Changed 9 years ago by ryanque

  • Summary changed from 05 Feb 2016 - VIPER (Ionosonde) and Digisonde (and synchronization) to 05 Feb 19:40 VIPER (Ionosonde) and Digisonde (and synchronization)
Note: See TracTickets for help on using tickets.