#4 new defect

Preventive Mainteinance 24/10/12

Reported by: lnavarro Owned by: somebody
Priority: major Milestone:
Component: component1 Version:
Keywords: Cc:
Parent Tickets:

Description

Problems with the Camera software sice 13/10/12. The Master software was stocked in the temperature loop. A test was performed using ANDOR software and the camera was not able to reach -78C, it was stocked in -80C.

The system will be shutdown and more test will be performed tonight.

Subtickets (add)

Change History (1)

comment:1 Changed 12 years ago by lnavarro

This problem was consulted with the main user(Dr. Meriwether) who indicated that this is due of the vaccum seal problem that this camera experienced before so a new temperature set point was set up in the scripts: -70C in the CameraControl? File instead of -78C.