#4 new defect
Preventive Mainteinance 24/10/12
Reported by: | lnavarro | Owned by: | somebody |
---|---|---|---|
Priority: | major | Milestone: | |
Component: | component1 | Version: | |
Keywords: | Cc: | ||
Parent Tickets: |
Description
Problems with the Camera software sice 13/10/12. The Master software was stocked in the temperature loop. A test was performed using ANDOR software and the camera was not able to reach -78C, it was stocked in -80C.
The system will be shutdown and more test will be performed tonight.
Subtickets (add)
Note: See
TracTickets for help on using
tickets.
This problem was consulted with the main user(Dr. Meriwether) who indicated that this is due of the vaccum seal problem that this camera experienced before so a new temperature set point was set up in the scripts: -70C in the CameraControl? File instead of -78C.